Abstract

In this work, design, manufacture and surgical success of a personalised temporomandibular prosthesis is featured. A fused deposition modelling technique and Die forging process constitute the methodology used in a patient who had an amputation in the upper third branch of the mandible, without considering the joint capsule. The implant was designed using a processed resection image of a computational tomography and using the methodology of Özkaya and Nordin. The jaw operating conditions were simulated by the finite element method (FEM). The main considered factors were the morphological geometry of the patient, implant fixation in the first third of the branch, implant fixation on the chin, dental post for placement of the teeth, and the form of the sub-lingual fossa weight optimisation. Special consideration was to preserve the patients facial aesthetics.

Keywords: CAD|CAM; Custom prosthesis; FDM; FEM; Digital temporomandibular models;

2 Introduction

In selecting the best method to reconstruct the low jaw there are important factors to be considered such as patient psychological aspects, the affecting disease, the expected quality of life, etc. The applied reconstruction method could be spongy cells on bone grafts, or bone reconstruction. Even with technological advances, complications could exist in jaw movement to chew, deficient sensing, infections at the treated zone, and weakness in the strength of the face structure. It is important to grant aesthetic aspects to the face in order to reinforce the natural person's confidence. Other aspects to be considered are age, height and weight of the jaw bone, as well as the state of teeth.

From open literature, [1] reported the biomedical comparison of two reconstructive methodologies applied to jaw fracture. The first one consists in applying a fixed Titanium plate, fixing it to the lateral edge of the jaw. The second method applies the same characteristics technique used by the first procedure, but fixed to the low jaw edge. Some significant differences were found in regard comparison to the implant mechanical behaviour. [2] proposes a method based on the manufacture of the a replacement with specific measures of the patient's jaw, obtained prior to surgical procedure; prosthesis manufacture was very close to the original bone configuration of the patient, which was taken before the surgery. [3] developed a 3D study, from tomographic resources, which allowed the existence of mineralisation of the cortical bone distribution part [3]. Additionally, [4] applied a 3 dimension (3D) model to known the mechanical behaviour of the jaw, determining that it bears a load up to 107N/m. This loading condition is generated by impact, and could lead to fractures at low jaw [4].

[5], developed an anosteosynthesis plate for an atrophic jaw, by using a mono cortical osteosynthesis of 2 Titanium plates. [6] compared the positions of implant subjection elements for grafts implants used into a jaw reconstruction. It concludes that a subjection element at the plate produces a better behaviour at the connection implant when chewing. [7] developed a study to determine the human jaw elastic response, by using finite element algorithms derived from mechanical loads. [8] used Finite Element Method (FEM), to simulate lateral movement of jaw, as well as collateral ligaments that connect articulate discs to condyle poles. This study determined that the function of temporomandibular ligament is to guide the jaw during its motion. At the same time, [9] developed a mandible reconstruction due to a tumour extraction in a Tanzanian patient. Vascular micro-grafts were used to reconstruct the atrophied jaw parts, concluding that bone grafts immediately inserted after resection could be better treated for these types of tumours.

[10] performed a complete jaw numerical analysis by FEM; the model construction was estimated by contour conditions, the loading settings and the jaw limit displacement circumstances. Likewise, [11] performed a jaw reconstruction by means of free human tissue transfer. The conclusion was that the main parts are iliac and scapula, which allow the mobility of the tongue, so they have to be considered in jaw reconstruction. [12] developed the numerical analysis of a 3D biomechanical model of the oral cavity to predict the consequences of tongue surgery on tongue movements, according to the size and location of the lost tissue, as well as the nature of the flap used by the surgeon. [13] analysed jaw reconstruction using 3D models of calcium in order to manufacture a physical model of jaw prosthesis; They consider that performing a pre-operational analysis, should allow more accurate reconstruction plates improving implant alignment.

With respect to patents, a wide range of geometries and morphologies for implants has been reported, that aid in the treatment of diverse pathologies. [14] refers to a modular type of mandibular prosthesis used for jaw resection, which considers the replacement of the jaw bone, by analysing the total jaw angle; prosthesis was developed by a sub-application of a couple of plates and a set of links used as connector elements. [15] proposes a bio-compatible prosthetic device for the replacement of temporomandibular joint, which contemplates the substitution of the condyle neck, the condyle, as well as condyle pit; the device includes a convex spherical surface condyle component of the first radio for the substitution of the articulate surface of the condyle head.

[16] develops a new fixation method, that allows a free rotatation of the elements in a portion at the edge that surrounds the fixation protuberance, over a bolted extreme neck that extends from the fixed plate to the mandibular. The plate has fixation slits to be anchored to jaw, allowing the adjustment in length, angle and shape of the mandible; it permits a much better performance. [17] shows a mandibular prosthesis design, for its usage when jaw bone resection of condyle. It includes four holding elements to the mandibular body, to maintain jaw angle's geometry; these are separated by a distance similar to the thickness of the jaw. [18] presents a prosthesis consisting of an adjustment screw aligned to the body of the jaw, which allows to modify its length. It is applicable in cases of orthognathic surgery, in order to align jaw and maxilla. The device is fixed to the mandible using infra-osseous plates and screws. The prosthesis angle can be modified between 90 to 150 degrees, which allows to consider patient's anatomy, based on tomographic studies. [19] presents a case of a total jaw implant, designed and manufactured using stereo-lithographic techniques. This prosthesis contains a hidroxyapatite layer and additives in the condyle zone. The mandibular body contains a series of spaces, to decrease the implant weight. This prosthesis allows guiding of the facial nerve through the configuration channel.

This work is focused on the development of customised design of implants used for the partial restitution of jaw applied to a Mexican patient of medium age, who underwent cancer. A multidisciplinary solution was developed, applying engineering mechanical tools, medical surgery advanced techniques, X-Ray tomography data, Finite Element Method simulations and a fused deposition modelling technique, in order to optimise a forging manufacturing process, to produce customised lower jaw prosthesis. This work is organised as follows: in Section 3 the clinical background as well as materials and methods used in this work, including the description of the numerical analyses (FEM) herein performed (Subsection 3.1), were provided. In section 4 the results are shown, while in Section 5 concluding remarks along with a profuse discussion on the results obtained is performed.

3 Background, methods and materials

As background, a male person around 40 years old came to an ordinary dentist appointment suffering from an apparent uncontrollable molar pain. The dentist treated the pain and took an X-Ray tomographic slice, where a bone tumour was found. The present research started with the X-Ray tomography study and the biopsy analysis. After an extensive evaluation, a bone cancer tumour was determined to be the cause of pain. The bone tumour had a cystic aspect, osteolytic, multi-ocular, with well-defined and clear edges, of approximately 6 cm long and 3 cm height and located at the left lower jaw size. The tumour was extracted surgically, by a mandibulectomy or jaw resection. It was necessary to place a Titanium chin support in order to avoid a collapse in the muscles and face tissue.

Thus, implant that could avoid deformation of the mandibular tissue, dropping of the face muscles, was required. It was necessary to prevent an misalignment of teeth so as to facilitate the daily activities (speech in a normal way, chewing, feeding, swallowing). This is a regular situation when facing a partial lost of the mandible bone due to the extraction of a cancer tumour.

Cross view of Computer Axial Tomography Reconstruction.\label{F2a} 3D model reconstruction.\label{F2b}
(a) Cross view of Computer Axial Tomography Reconstruction.\label{F2a} (b) 3D model reconstruction.\label{F2b}
Figure 1: X-Ray process.

X-Ray tomography data allowed damage evaluation from surgical procedure, and determined the volumetric space vacancy at the lower jaw (see Figure F2a). On the other hand, ScanIp software was used to process 400 slices, so to develop patient's head. Figure F2b shows jaw resection and a simple chin that maintains the aesthetics and functionality of the remaining third branch of the jaw.

Clinical studies of the patient were used to generate a personalised implant proposal. A review of jaw prosthesis available in the domestic and international markets allowed to establish some important parameters to be considered for the customised design of the jaw endo-prosthesis. The most important ones are:

  1. Jaw prosthesis anatomy, which had to be adapted to the biomechanics of the joint (Zygomatic bone diameter, jaw bone transverse and longitudinal section).
  2. Jaw component orientation.
  3. Variations in the cross section of the central third jaw bone.
  4. Anchorage surface between jaw bone and joint.
  5. The necessity of no displacement between jaw bone and joint.
  6. Determination of static, dynamic and fatigue loads by analysing the jaw prosthesis curvature.
  7. Anti-rotation effect.
  8. Determination of the physical conditions of the receptor bone, considering permanent fixation.
  9. Material selection by physical and chemical characteristics to allow structural integrity and minimum wear.
  10. Natural external agent distribution into the jaw bone and in to the prosthesis.

It should be pointed out that jaw prosthesis service life is directly related to displacement, micro-movements and implant rotation. Thus, prostheses ought to have smooth surfaces and avoid sharp edges, in order to eliminate stress concentration into the implant and in the elements that should be used to fix the jaw prosthesis at the mandibular joint. Jaw prosthesis must be wider in the lateral direction, far away from the average midline face; it should not be very rigid, have enough stability to control vibrations and should anticipate particle migration from the articulate surface towards the zygomatic area. Thus, this work allows the establishment of a criterion for the development of a personalised jaw prosthesis geometry. The initial geometry for the design of the implant is taken from X-Ray tomography data, to generate the working planes necessary for each mandibular bone area, as it is shown in Figure F2a. An exterior silhouette of the main mandibular geometry was developed, from which it was possible to obtain the bold dimensions of the implant, the main angles of the jaw and the internal relations of the condylar alignment. In the same way, a line is drawn that crosses through the centre of all the circumferences that wrap the jaw healthy body, so as to obtain the average curvature line and establish the maximum and minimum values for the teeth pieces alignment.

Mandibular resection.\label{F3a} Reconstruction sagittal view.\label{F3b} Inferior view of the mandibular bone alignment.\label{F3c}
(a) Mandibular resection.\label{F3a} (b) Reconstruction sagittal view.\label{F3b} (c) Inferior view of the mandibular bone alignment.\label{F3c}
Figure 2: Divers view of the face-jaw model reconstruction.
3D Model of the patient.\label{F4a} Partial jaw model.\label{F4b} Final assembly of the implant.\label{F4c}
(a) 3D Model of the patient.\label{F4a} (b) Partial jaw model.\label{F4b} (c) Final assembly of the implant.\label{F4c}
Figure 3

In Figure F3a, a general view of the main model for the patient face deficiencies without the chin, is shown. This model is extremely important because the accuracy is a variable that delimitates the success of the patient's aesthetic. As mentioned by [20], the manufacture by means of stereolithography shows some significant differences, but these are clinically irrelevant; for this reason, fused deposition modelling was the used technique to evaluate the behaviour and assembly of mandibular endo-prosthesis (also see Figure 3). Figure F3b shows a sagittal view, where the simulation of the implant that will be placed at the mandibular joint is featured. In Figure F3c, an inferior view of the half jaw replacement can be appreciated, which also presents a better perspective of the new jaw prosthesis. This constitutes the model that will be used for FEM analyses. The model of the implant was built up with Powersheapesoftware, which allowed the development a half mirror model of the missing jaw. It also allowed the construction of anchored zones for customised implant and the alignment of posts for the placement of teeth.

The jaw prosthesis was modelled with an ABS polymer before the reconstruction surgical procedure, in order to check the elements assembling. This polymer model allowed the simulation of one of the three-dimensional available designs of the jaw prosthesis. The 3D face bone model allowed the simulation of the face of the patient, so to identify areas that surgeon should cut off as well as space voids that should be filled up, thus, determining the best way to insert prosthesis. Additionally, useful information from the equipment to be used in the surgical procedure, consists in data on the contact that should be generated to improve the union between the implant against the joint bone. In this way, it is possible diagnose a treatment plan. This can only be possible by obtaining the anatomical model of the damaged area with X-Ray tomographic shots.

It is important to mention, that for every clinical case, it is possible to design and develop a partial or complete prosthesis with this methodology. The later depends on patient, his pathology, patient's age and the medical surgeon indications in order to establish the best technology for the substitution of the bone structure that has been damaged. Nonetheless this advantages, our proposal results in a handcrafted procedure. Figure 3 shows the use of fused deposition modelling technique to a 3D model.

\label{F5a} \label{F5b}
(a) \label{F5a} (b) \label{F5b}
Figure 4: Different views of the developed lower jaw prosthesis.

The new lower jaw implant shows a prosthesis device that should allow the patient to maintain his teeth, bite, eat, swallow, as well as a correct alignment. Figure 4 shows jaw prosthesis. This device was generated from computer X-Ray tomography taken from the patient, which helped to determine specific and personal data, so to generate the optimised solution for the person. The principal advantage is the obtainment of customised prosthesis that considers the patient's morphological geometry, as well as a specific suitable fixation between condyles and chin.

3.1 Numerical analysis

A numerical model is used to perform numerical analyses, for which is necessary to know the mandibular anatomic characteristics. Getting the model through X-Ray scans, simplifies and reduces time to generate the numerical simulation. In order to generate it, a simple transformation of scanned data is required, in order for them to be able to be used in FEM. In order to apply the later method, boundary conditions due to external agents ought to be determined. The loads that mouth muscles can exert during daily activities, can be estimated from two theories [21,22,23,24,25]:

  1. The exerted load by the mouth muscle when contraction is proportional to its transversal section.
  2. The muscle contraction varies depending on the time of the activity.

So, the resulting load that the mouth muscle is able to apply during its contraction is described as ():

(1)

where: refers to a physiologic constant, refers to the muscle cross section, and refers to the activity. The value is considered as 40 N/cm [22], independently of the age, sex and muscle of the person. There are studies where cross section of the muscle has been measured [22]. Table 1 shows average obtained values.


Table. 1 Average values of chewing muscles [22].
Muscular group Section (cm) Muscle portion Size portion Section (cm) (N)
Masseter 6.80 Superficial 0.70 4.76 190.4
Depth 0.30 2.04 81.6
Pterigoidmedium 4.37 1.00 4.37 174.8
Temporal 8.23 Anterior 0.48 3.95 158.0
Medium 0.29 2.39 95.6
Posterior 0.23 1.89 75.6
Pterigoid lateral 2.39 Inferior 0.70 1.67 66.9
Medium 0.30 0.72 28.70
Digastric 1.00 1.00 1.00 40.0

The total average chewing load for all muscles is 911.6 N. The maximum anatomical chewing load corresponds to the maximum contractile power of the muscles when they open the jaw. The chewing load is calculated based on the relation between the load developed by a skeletal muscle and the muscular cross unit area. [26] reported that chewing load is between 210 to 400 kg (3924 N). The numerical analysis presented in this work considered diverse parameters such as:

Average maximum loads, i.e. that the load is uniformly distributed along the chewing line. The numerical simulation was performed in a 3D model, with high order elements with 20 nodes. The analysis considers applied elastic, isotropic and homogeneous conditions. Figure 5 shows applied boundary conditions at the edges of jaw implant, as well as the exerted loads. The mechanical properties used are those for Titanium (ELI) Grade 5 alloy alpha beta with 6% Al and 4% V, Young’s modulus 200 GPa and Poisson relation of 0.28.

Loading and boundary conditions at the jaw prosthesis.
Figure 5: Loading and boundary conditions at the jaw prosthesis.

von Misses stress and equivalent strain with an applied average load of 911.6 were calculated. Figure 6 show total displacement, von Misses stress and equivalent strain of the implant when undergoing a maximum load of 3924 N.

Total displacement.\label{F7a} von Misses stress.\label{F7b} Strain.\label{F7c}
(a) Total displacement.\label{F7a} (b) von Misses stress.\label{F7b} (c) Strain.\label{F7c}
Figure 6: Maximum loading results.

4 Results

The half-lower jaw prosthesis, showed in Figure 4, was manufactured with a biocompatible material (Titanium ELI) using only one piece and applying compression forces during the manufacturing process. The implant was modelled to provide a similar morphological pattern, which allow the optimal replacement of the mandibular resected bone. The endo-prosthesis has a series of holes to decrease prosthesis weight, without decreasing its structural integrity. One of its ends has the form of the mandibular notch and neck, allowing its connection to the condyle through an anchorage, that holds the condyle to the upper jaw. It is possible to apply this kind of attachment because the jaw angle depends on the morphological characteristics of the patient. Both ends of the prosthesis are screwed to the patient with surgical bolts, as it is shown in the Figure 7.

New half lower jaw prosthesis presented into the model.
Figure 7: New half lower jaw prosthesis presented into the model.

Results from the numerical simulation show that von Misses stresses, strains and total displacements are low when an average chewing load is applied. These are under the mechanical risk zone that Titanium could undertake, so the implant should work properly under normal chewing conditions. If the chewing load reaches a maximum value, the implant should work just under its mechanical yield zone; so if the patient is exposed to this condition it would not reach a critical value. Nevertheless, several physiological and physical factors would not permit to reach the maximum load.

5 Conclusions

In this work, a personalised half lower jaw prosthesis using X-Ray tomography and a fused deposition modelling technique, was developed. The prosthesis was successfully implanted in Mexico, to a patient suffering jaw bone cancer, as it is shown in Figure 8. The surgery procedure, performed to the patient, was to partially amputate the mandibular bone. The innovation presented in this work, is the application of 3D technology (modelling) to reconstruct half of the lower mandibular bone by inserting an endo-prosthesis.

As mentioned by [27], computer-assisted procedures have proven to be effective methods to obtain dental alignment between the upper and lower jaw. Nevertheless, as in this work the developed prosthesis was customised, it naturally provides a correct alignment with the face structure and with the healthy mandibular bone. Additionally, prosthetic teeth can be attached to tolerate normal chewing. Also, the customisation of the prosthesis should allow a better life quality to the patient. The development of this prosthesis can be used to relieve three different mandibular bone deficiencies; as an artificial condyle, as a band for the substitution of the principal jaw bone and as a strip band for additional fixation member.

Frontal plane.\label{F9a} Transverse plane.\label{F9b}
(a) Frontal plane.\label{F9a} (b) Transverse plane.\label{F9b}
Figure 8: Jaw surgery.

Consequences of deceases are diverse, and might lead to total or partial loss of a member by amputation, that can significantly affect the patient. The replacement of an organ or a member with prosthesis can represent, for the person, a recovery of an aptitude to perform different tasks and even organic basic functions. According to [28], psychosocial aspects toughen the patient in the recovery phase from surgery, mainly do to the number of successful - or not - quirurgical procedures. Such psychosocial deterioration is sometimes more incapacitating than the perception of the efficiency of undertaken surgical approach. Besides psycosocial, aesthetic aspects also affect the patient. All these factors contribute to his quality of life.

For the case of an injury in the head or face, this involves an important health risk. The surgical amputation would have lifetime consequences for the patient. The first stage, once the patient has recovered from surgical medic process (organ stabilisation), consists on evaluating and attending the functional and aesthetic consequences of the procedure. In the case of the lower jaw loss, the immediate consequences are related to the respiratory system, the digestive system, the vocal system and the evolution of organ reactions depending on sleeping. The prosthesis restitution implies, first of all, the recovery of organ systems and basic functions, from the medical point of view. From the patient's point of view, it is important to restore a natural aesthetic appearance, which could bring confidence to the person. Within this first stage, there is a consensual solution for both parts (medical and patient): The prosthetic implant must adjust to the biological appearance and perform the required functions. For the new low jaw prosthesis, developed in this paper, it is aimed to provide both.

In the second stage of recovery, the manufactured prosthetis should be refined and detailed for the customisation for the patient and his requirements, like it is considered by [29]. They perform three-dimensional soft tissue modelling of chin, for implementing an aesthetics treatment. For this case, the prosthetic implant has already been personalised to suit morphological aspects of the patient. So, it is expected that the prosthesis integration occurs in a natural manner to the respiratory and digestive biological processes. Additionally, it should no difficult its integration of adjacent organs (tongue, teeth, trachea, chew, etc.). Within the aesthetic aspect, the general integration should arise in patient's mind. The total recovery after surgery should provide confidence to the patient and his face should match or be very similar to the one before surgery, conserving his face symmetry, without cosmetics factors, and his functions properly working.

In a third and last recovery stage, the application and normal service (rehabilitation) of the prosthesis was assessed. Prosthesis should be analysed and corrected, in case of any collateral effects, organ adjustments (life of the prosthetic), as well as not working properly, inherently. Cosmetic variations must be determined (ageing in comparison to prosthetic), which could require adjustment interventions. These should be easy, due to the prosthesis design herein developed.

5.1 Discussion

This research deals with a system of metal mandibular prosthesis for patients who have undergone a subtotal resection hemimandibulectomy. This prosthesis was developed in order to restore damaged functions when part of the jaw was injured by tumour cancer. A metal mandibular prosthesis system is developed to provide consolidation, hygiene and easiness in bone implant, and it is able of respond to a partial resection without affecting the jaw or temporomandibular joint condyle, the replacement of teeth, having mobility axes: X, Y and Z rotatory capacity, simulating natural bite. It also provides aesthetic volume, turn the volume and tone facial muscles in order to provide an adaptation to any patient's morphological geometry. Although designed for humans, the procedure herein developed to design and manufacture jaw prosthesis could be easily extended for its usage in veterinary area.

Acknowledgements:

The authors gratefully acknowledge the financial support from the Mexican government by Consejo Nacional de Ciencia y Tecnología, to the Instituto Politécnico Nacional and the biomechanics group of INSIA from the Universidad Politécnica of Madrid. Authors also acknowledge partial support projects 20181139, 20180472, 20181441, 20181028, 20181141, as well as EDI grant, all provided by SIP/IPN.

Conflict of Interest

The authors declare that there is no conflict of interest regarding the publication of this paper. The founding sponsors had no role in the design of the study; in the collection, analyses, or interpretation of data; in the writing of the manuscript, and in the decision to publish the results.

BIBLIOGRAPHY

[1] Madsen, Matthew J and Haug, Richard H. (2006) "A biomechanical comparison of 2 techniques for reconstructing atrophic edentulous mandible fractures", Volume 64. Elsevier. Journal of oral and maxillofacial surgery 3 457–465

[2] Tideman, H and Samman, N and Cheung, LK. (1998) "Functional reconstruction of the mandible: a modified titanium mesh system", Volume 27. Elsevier. International journal of oral and maxillofacial surgery 5 339–345

[3] Maki, K and Miller, AJ and Okano, T and Shibasaki, Y. (2001) "A three-dimensional, quantitative computed tomographic study of changes in distribution of bone mineralization in the developing human mandible", Volume 46. Elsevier. Archives of Oral biology 7 667–678

[4] Torreira, Mercedes GALLAS and Fernandez, Jose Ramon. (2004) "A three-dimensional computer model of the human mandible in two simulated standard trauma situations", Volume 32. Elsevier. Journal of Cranio-Maxillofacial Surgery 5 303–307

[5] Seper, László and Piffkó, József and Joos, Ulrich and Meyer, Ulrich. (2004) "Treatment of fractures of the atrophic mandible in the elderly", Volume 52. Wiley Online Library. Journal of the American Geriatrics Society 9 1583–1584

[6] Nagasao, Tomohisa and Miyamoto, Junpei and Tamaki, Tamotsu and Kawana, Hiromasa. (2010) "A comparison of stresses in implantation for grafted and plate-and-screw mandible reconstruction", Volume 109. Elsevier. Oral Surgery, Oral Medicine, Oral Pathology, Oral Radiology, and Endodontology 3 346–356

[7] Clason, Christian and Hinz, Andreas M and Schieferstein, Heinrich. (2004) "A method for material parameter determination for the human mandible based on simulation and experiment", Volume 7. Taylor & Francis. Computer methods in biomechanics and biomedical engineering 5 265–276

[8] Del Palomar, A Perez and Doblaré, M. (2006) "Finite element analysis of the temporomandibular joint during lateral excursions of the mandible", Volume 39. Elsevier. Journal of Biomechanics 12 2153–2163

[9] Simon, ENM and Merkx, MAW and Shubi, FM and Kalyanyama, BM and Stoelinga, PJW. (2006) "Reconstruction of the mandible after ablative surgery for the treatment of aggressive, benign odontogenic tumours in Tanzania: a preliminary study", Volume 35. Elsevier. International journal of oral and maxillofacial surgery 5 421–426

[10] Liao, Sheng-Hui and Tong, Ruo-Feng and Dong, Jin-Xiang. (2007) "Anisotropic finite element modeling for patient-specific mandible", Volume 88. Elsevier. Computer methods and programs in biomedicine 3 197–209

[11] Bak, Mathew and Jacobson, Adam S and Buchbinder, Daniel and Urken, Mark L. (2010) "Contemporary reconstruction of the mandible", Volume 46. Elsevier. Oral oncology 2 71–76

[12] Buchaillard, Stéphanie and Brix, Muriel and Perrier, Pascal and Payan, Yohan. (2007) "Simulations of the consequences of tongue surgery on tongue mobility: implications for speech production in post-surgery conditions", Volume 3. Wiley Online Library. The International Journal of Medical Robotics and Computer Assisted Surgery 3 252–261

[13] Katsuragi, Yoko and Kayano, Shuji and Akazawa, Satoshi and Nagamatsu, Shogo and Koizumi, Takuya and Matsui, Takahiro and Onitsuka, Tetsuro and Yurikusa, Takashi and Huang, Wei-Chao and Nakagawa, Masahiro. (2011) "Mandible reconstruction using the calcium-sulphate three-dimensional model and rubber stick: a new method,‘mould technique’, for more accurate, efficient and simplified fabrication", Volume 64. Elsevier. Journal of Plastic, Reconstructive & Aesthetic Surgery 5 614–622

[14] Manolidis, Spiros. (2000) "Modular mandibular prosthesis". Google Patents

[15] Gordon, Jeffrey D. (1996) "Apparatus for total temporomandibular joint replacement". Google Patents

[16] Nealis, Michael F. (1995) "Temporomandibular joint prosthesis". Google Patents

[17] Collins, Thomas A. (1988) "Mandibular prosthesis". Google Patents

[18] Touleimat, Saad and Touleimat, Abdul-Elah. (2014) "Adjustable mandible joint system". Google Patents

[19] Van Loon, Jan-paul. (1999) "Adjustable temporomandibular surgical implant". Google Patents

[20] Camardella, Leonardo Tavares and Vilella, Oswaldo V and van Hezel, Marleen M and Breuning, Karel H. (2017) "Accuracy of stereolithographically printed digital models compared to plaster models". Springer. Journal of Orofacial Orthopedics/Fortschritte der Kieferorthopädie 1–9

[21] Pruim, GJ and De Jongh, HJ and Ten Bosch, JJ. (1980) "Forces acting on the mandible during bilateral static bite at different bite force levels", Volume 13. Elsevier. Journal of biomechanics 9 755–763

[22] Weijs, WA and Hillen, B. (1984) "Relationship between the physiological cross-section of the human jaw muscles and their cross-sectional area in computer tomograms", Volume 118. Karger Publishers. Cells Tissues Organs 3 129–138

[23] J.W.C. MacDonald and A.G. Hannam. (1984) "Relationship between occlusal contacts and jaw-closing muscle activity during tooth clenching: Part I", Volume 52. The Journal of Prosthetic Dentistry 5 718 - 729

[24] J.W.C. MacDonald and A.G. Hannam. (1984) "Relationship between occlusal contacts and jaw-closing muscle activity during tooth clenching: Part II", Volume 52. The Journal of Prosthetic Dentistry 6 862 - 867

[25] E. Moller. (1996) "The chewing apparatus; An electomyographic study of the action of the muscles of mastication and its correlation to facial morphology", Volume 280. Acta Physiological Scandinavian Supplement 1-229

[26] G. Nelson. (1986) "Three dimensional Computer modeling of human mandibular biomechanics". The university of British Columbia 201-226

[27] Zizelmann, C and Bucher, P and Rohner, D and Gellrich, N-C and Kokemueller, H and Hammer, B. (2010) "Virtual restoration of anatomic jaw relationship to obtain a precise 3D model for total joint prosthesis construction for treatment of TMJ ankylosis with open bite", Volume 39. Elsevier. International journal of oral and maxillofacial surgery 10 1012–1015

[28] Guarda-Nardini, L and Manfredini, D and Ferronato, G. (2008) "Temporomandibular joint total replacement prosthesis: current knowledge and considerations for the future", Volume 37. Elsevier. International journal of oral and maxillofacial surgery 2 103–110

[29] Holberg, Christof and Heine, Anne-Kathrin and Geis, Philipp and Schwenzer, Katja and Rudzki-Janson, Ingrid. (2005) "Three-dimensional soft tissue prediction using finite elements", Volume 66. Springer. Journal of Orofacial Orthopedics/Fortschritte der Kieferorthopädie 2 122–134

Back to Top

Document information

Published on 04/03/19
Accepted on 25/02/19
Submitted on 13/02/18

Volume 35, Issue 1, 2019
DOI: 10.23967/j.rimni.2019.02.001
Licence: CC BY-NC-SA license

Document Score

0

Views 290
Recommendations 0

Share this document